cannon tesc全自动低温布氏粘度测定仪 astm d2983
cannon tesc全自动低温布氏粘度测定仪 astm d2983 产品详情
品名:tesc全自动低温布氏粘度测定仪 astm d2983
型号:tesc
品牌:cannon
应用:粘度测试
类别:粘度测定仪
适用标准:astm d2983
产品特点:全自动;
cannon
tesc thermoelectric sample conditioner system for astm d 2983
tesc全自动低温布氏粘度测定仪
符合astm d2983
自动化温度调节和测试,控温范围:-40℃-+90℃
允许样品在整个预热、室温下稳定、冷却和终粘度测试保持恒定
避免温度波动导致的结果变异
整个温度调节过程比传统的d2983精度更高
全自动减少仪器设置时间
体积小巧,可安装多个tesc系统在一个相对较少的台面
可以方便的调节粘度计转子
半导体制冷技术
控温精度±0.1
ordering information
description
part #
thermoelectric sample conditioner
(tesc) system: includes the brookfield® dv2t viscometer and support, temperature control software, #4b2 spindle, ez-lock spindle coupling, pre-loaded test programs on usb memory stick, usb cable, test cells (12 pack), n27c viscosity reference standard (500 ml) and rheocalc software. the unit is pre-calibrated for testing gear oil at -40 °c.
9725-f81
accessories & consumables
description
part #
test cells (case of 72); 25 mm od x 150 mm, imless,disposable
9725- f83
test cells (pkg of 12); 25 mm od x 150 mm, rimless, disposable
9725-f84
viscosity reference standard n27c
(viscosity in cp at –40 °f, –30 °f, –20 °f, –15 °f, –10 °f
and 0 °f)
9727-g12.016
viscosity reference standard n115b
(viscosity in cp at –20 °f, –15 °f, –10 °f, 0 °f, +10 °f
and +20 °f)
9727-g15.016
viscosity reference standard n14b
(17,000 cp at –40 °c)
9727-g65.016
viscosity reference standard n120b
(150,000 cp at –40 °c)
9727-g30.016
viscosity reference standard n480b
(150,000 cp at –26 °c)
9727-g35.016
viscosity reference standard n1400b
(150,000 cp at –12 °c)
9727-g40.016
brookfield® #4b2 insulated spindle (replacement)
17.5128
brookfield® ez-lock spindle coupling (replacement)
17.5129
product specifications
dimensions (w x d x h)
16.5 cm x 51 cm x 76 cm;
6.5 in x 20 in x 30 in (including dv2t viscometer)
weight
16.3 kg; 36 lb (including dv2t viscometer)
temperature range & accuracy
–40 °c to +90 °c (± 0.1 °c)
sample capacity
1
computer requirement
a dedicated pc is not required for standard instrument
operation. a pc is necessary if reprogramming for tests other than gear oil at -40 °c
power requirements
tesc unit: 100/240 vac, 50/60 hz, 320 watts
dv2t viscometer: 90/260 vac, 50/60 hz
operating conditions
15 °c to 30 °c, 10% to 90% relative humidity (non-
condensing), installation category ii, pollution degree 2
compliance
ce mark; emc directive (2004/108/ec); low voltage
directive (2006/95/ec); rohs