- 加工定制:
- 品牌:Gold APP Instruments
- 型号:H-Sorb 2600
- 类型:高压气体吸附仪
- 品种:其他
- 测量范围:无
- 准确度:0.1%
- 显示方式:数字
- 分度值:
- 电源:220
- 用途:实验室
agentsare wanted
contactinfo:
web:
tel.: 0086-10-82132122
fax:0086-10-82118197
mobile:0086-
ims:
skype:gold-app-instruments
yahoomessenger:
h-sorb 2600 high pressure gas adsorption volumetric principleanalyzer introduction
gas sorptionanalysis is important in many fields of materials, such as energystorage materials, improved catalysts for petrochemical processing,advanced pharmaceuticals and food industries. at present, there istremendous interest in the development of innovative materials forthe transition to renewable energy. the science of gas sorption hasbecome particularly critical in the advancement of materials forstoring fuel gases such as hydrogen or natural gas, as well as forsequestration of greenhouse gases.
the h-sorb 2600 high pressure gas sorption analyzer adoptsthe volumetric technique principle which consists of dosing a knownamount of gas adsorptive into measurement tubes where lying samples(such as zeolites, carbon nanotubes, rocks, catalysts, activatedcarbons, molecular sieve, hydrides etc. ) to be analyzed. whensamples reach equilibrium with the adsorbate gas (such as hydrogen,methane, carbon dioxide, nitrogen, argon, krypton etc.), detailedand comprehensive final equilibrium pressures will be recordedautomated by gold app instruments company dedicated and patentedsoftware. these data are used to calculate the quantity of gasadsorbed by samples. this measurement procedure automaticallyrepeated by software under given temperature and pressure intervalsuntil finalized the maximum preselected pressure. finally, each ofthe equilibrium points’ volume adsorbed, equilibrium pressure,absolute pressure etc. data are plotted to produce the isotherm. byusing the isotherm data, operators can calculate further moreinformation such as langmuir fitting isotherm, the adsorption ratecurve, pct (pressure-composition-temperature) or pci(pressure-composition-isotherm) curve etc.
pressure-composition isotherms (pci) andpressure-composition temperature (pct) is one of the mostinformative sorption measurements. the result is a plot of theequilibrium absorbed gas concentration in the material as afunction of pressure and temperature.
gold app instruments high pressure and high temperature gassorption analyzer h-sorb 2600 adopts static volumetric principle todetermine adsorption and desorption isotherms under high pressure(atmospheric to 200bar) and high temperature (atmospheric to500℃) with astainless steel vcr manifolds system, fully automated operation cancontrol system and multi calculation methods, all these guaranteeh-sorb 2600 to produce a highly repeatibility and highly precisionanalysis data.
h-sorb 2600 high pressure gas adsorption volumetric principleanalyzer application
study and research in high temperature and pressure gassorption, supercritical gas performance, microporous materials,hydrogen storage materials, coalbed gas, oil exploration, methaneetc.
catalysts, molecular sieves, activecarbon, carbon nano-tube, fuel cell, hydrocarbon and othermaterials.
h-sorb 2600 high pressure gas adsorption volumetric principleanalyzer parameters
analysismethod: staticvolumetric principle, high pressure gas sorptionprinciple
versatility: adsorption/desorption isotherms measurement (atmospheric to500℃),gibbs supercritical adsorption measurement
datareduction: langmuirmodel regression isotherm, pct(pressure-composition-temperature) orpci (pressure-composition-isotherm) diagram/curve/isothermmeasurement, langmuir maximum adsorption constant l and adsorptionpressure constant b measurement; langmuir correction modelloading-ratio correlation regression (lcr) isotherm; ono-kondo (ok)lattice model isotherm regression; adsorptive gas density andvapor-liquid equilibrium accuracy calculation under differentpressures and temperatures
measuringranges: continuousadsorption and desorption measurement from atmospheric to 200barpressure
accuracy: repeatability errors <3%
temperaturerange: cryogenicto 500℃,accuracy±0.1℃
sampleports: two samplesanalyzing and two degassing, stainless-steel micro welding samplecells, software integrated temperature pid adjustment
transduceraccuracy: importedhigh precision transmitter, accuracy can reach 0.05% of f.s.(full-scale), long-duration usage stability is 0.025% off.s.
ultimatevacuum: 4x10-2 pa (3x10-4 torr), molecularpump system (1x10-6 pa) is optional
adsorbategas: high purity nitrogen, co2or others (ar, kr, h2, ch4 etc.)
dataacquisition: high-precision and high integration data acquisitionmodules, minimal error, strong anti-interference ability
specifications: height22.05 inches (56cm)* width 20.87 inches (53cm)*depth 23.62 inches(60cm); 132 pound (60kg); 220v/5a
h-sorb 2600 high pressure gas adsorption volumetric principleanalyzer features:
control system: imported vcr interfacehigh pressure pneumatic valve can realize auto on-off within200bar, sealing performance is top to 1x10-10 pa.m3/s,can be used more than 500 million times; programmable logiccontroller (plc) system, high integration and soundanti-interference, improve stability and prolong life
analysismanagement: parameters be set by gold appinstruments software, fully automated, achieved unattendedoperation at night; h-sorb mode is available for accurate controlgas charging pressure points, gain ideal data points
manifoldsystem: imported 316l stainless steelthick-walled manifolds, micro welding sealing can reduce dead spacelargely; metal vcr connection is convenient for installation anduninstallation
safetymeasures: h-sorb asymptotic technology realizesauto gas-charging and degassing, prevents high pressure dangerscaused by artificial misoperation, and also minimizes largedifferential pressure’s impact to pressure transducer
our policy of continuous development may cause theinformation and specifications contained herein to change withoutnotice or liability.
北京金埃谱科技有限公司(总部)
Andy
18210009838